Sterilizers
A sterilizer can achieve the destruction of all pathogenic organisms using heat, pressure or radiation. International hygiene guidelines recommend using autoclaves, which perform sterilization of medical devices and instruments with saturated steam. Autoclaves are widely used in dentistry, medicine, podiatry, tattooing and veterinary medicine. The units for steam sterilization vary in size and function depending on the instruments intented to be sterilized..
Hospital Items which can be sterilized in steam steriliser:
- Stainless steel
- Metallic medical tools, surgical Instruments
- Contaminated solid items
- Plastic pipette tips (inside appropriate biohazard beds)
- Polypropylene (PP secondary containers) and polycarbonate (PC) plastics (keep containers and bags open so that the steam can penetrate them)
- Biological tissue culture flasks and plates
- Media Solutions
- Glassware (Pyrex, Pyrex type materials, or type I borosilicate glass)
- Paper (if placed inside a biohazardous autoclave waste bag)
- Stainless steel
- Latex gloves, vinyl (inside biohazardous autoclave waste bags)
- Hospital linens, textile materials
Why hospitals should use imported good quality steam sterilizers only ?
- Only few imported sterilisers can give uniform temperature with proper steam penetration
- To Check Steam steriliser quality Class 2 indicator should be used .A commercially available Bowie-Dick-type test sheet should be placed in the centre of the pack. The test pack should be placed horizontally in the front, bottom section of the sterilizer rack, near the door and over the drain, in an otherwise empty chamber and run at 134°C for 3.5 minutes. The test is used each day the vacuum-type steam sterilizer is used, before the first processed load
- Local sterilisers take 5 to 7 mins to pass the test and indicator colour not changed uniformly.
Imported Sterilizer

Why Melag ?
- Company Headquarter at Berlin, Germany
- Started in year 1951.
- Worldwide sales of over 5,00,000 steam sterilizers and autoclaves are proof of the international success and appreciation of the MELAG product line.
- More than 500 employees.
- Wide range of varieties available.
- EC certified.
- EN ISO 13485 Certified
- MDD Annex II excl. 4
MELAG Certifications

Melag Pro Limelag Line Model – Vacuclave 123

Features of Proline Vacuclave 123
- Powerful Class-B Vacuum Procedure – Reliable Sterilization of up to 6 Kg on 5 Trays in only 15 to 28 minutes ( Plus Drying )
- Smart Touch display with LED status – Touchscreen with intuitive interface for simple operation like your smartphone
- Connectivity & data logging – Compliance with legal requirement by saving the sterilization parameters via USB or Network.
- Stand Alone device concept – Integrated water tanks and highly effective air cooling for site independent operation without cooling water.
- Power save energy saving mode – Heater & display go into standby mode between sterilizations for top energy efficiency.
- Easy care routine concept – User friendly optimizations for efficient filter replacement, tank cleaning & water supply
Pro-Class Series

Features of Pro-Class Series
- Performance: Convincing performance: Perfect sterilization and drying results of up to 7 kg in only 20 – 39 minutes (Quick-Programs incl. prevacuum and drying).
- Safety: Documentation made easy: The high-precision microprocessor control monitors all the relevant process parameters and saves the logs on aligned documentation media.
- User-friendliness: The tried-and-tested MELAG 4-key concept in connection with digital tools makes for ease of use and avoids application errors in everyday use.
- Flexibility: As versatile as your requirements: Match the sterilization process using a fixed water connection or stand-alone autoclave and three chamber depths to suit your individual needs

Device Design | Model | Volume |
---|---|---|
Tap – Water Connection | Vacuklav 30 B+ Vacuklave 24 B+ Vacuklave 24 BL+ | 17 Ltr. 23 Ltr. 29 Ltr. |
Stand – alone | Vacuklay 31 B+ Vacuklave 23 B+ | 17 Ltr. 23 Ltr. |
Premium Class – Sterilizers
Model Name | Vacuclav 40 B+ | Vacuclav 41 B+ | Vacuclav 43 B+ | Vacuclav 44 B+ |
---|---|---|---|---|
Chamber Size | 25 CM X 35 CM Long | 25 CM X 35 CM Long | 25 CM X 45 CM Long | 25 CM X 45 CM Long |
Volume | 18 Ltrs | 18 Ltrs | 23 Ltrs | 23 Ltrs |
Load Quantity | Max 9 Kg Instruments, 2 Kg Textiles | Max 9 Kg Instruments, 2 Kg Textiles | Max 9 Kg Instruments, 2.5 Kg Textiles | Max 9 Kg Instruments, 2.5 Kg Textiles |
Dimensions (WxHxD) | 46 X 50 X 55 CM | 46 X 50 X 55 CM | 46 X 50 X 55 CM | 46 X 50 X 55 CM |
Weight | 55 Kg | 60 Kg | 69 Kg | 64 Kg |
Power Supply | 220 – 240 V / 50-60 HZ (3.4 Kw) |
Pro Class – Sterilizers